Standard low temperature incubators are intended for Biological Oxygen Demand (B.O.D.) applications. B.O.D. incubators are designed for wastewater treatment, not Drosophila incubation and growth. B.O.D. applications are not adversely affected by temperature fluctuations associated with defrost cycling to prevent freezing coils. Fruit Flies cannot tolerate such dramatic temperature change. Additionally, Fruit Fly normal metabolic processes produce a gas that contaminates the frost on the refrigerated incubator coils. When the coils are defrosted, a corrosive substance forms, leading to premature coil deterioration.
The Shel Lab Drosophila-Specific incubators function within the range of temperatures preferred by fruit flies. Rather than striving to maintain a distinct set point within 0.25°C, the incubators gently fluctuate between 17°C and 19°C.
Specifications |
Solid Door 20.3 cu. ft. | View Pane Door 20.3 cu. ft. |
Model | LIFLY | LIFLY-VIEW |
Capacity | 20.3 cu. ft. (574 L) | 20.3 cu. ft. (574 L) |
Weight | 246.00 lbs | 246.00 lbs |
Shelving | 8 supplied | 8 supplied |
Interior Outlet | 115V-1/230V-2 | 115V-1/230V-2 |
Temperature Range | 0°C to 29°C | 0°C to 29°C |
Ext. Dimensions WxDxH in.(cm) | 34.5x34.5x77.5 (87.7x87.7x196.9) | 34.5x34.5x77.5 (87.7x87.7x196.9) |
Int. Dimensions WxDxH in.(cm) | 27.0x23.0x56.5 (68.5x58.4x143.5) | 27.0x23.0x56.5 (68.5x58.4x143.5) |
Nominal Voltage | 115/230 | 115/230 |
Max. Energy Consumption (W) | 960/1140 | 960/1140 |